Instrument
Capacity
Operation
Size
Weight
Power
Operating range
|
Embryo Monitoring System with Time-lapse image capture
Six disposable EmbryoSlide™ trays holding 12 embryos each
Individual slides may be inserted and removed independently
W × D × H (60.3 × 56.0 × 43.5) cm / (23.7 × 22 × 17.2) in
55 kg / 121 lbs
110 or 220V
20-30oC
|
Image acquisition
Built-in microscope
Resolution
Illumination for image acquisition
Time between acquisitions
|
Computer-assisted detection and analysis of cell divisions
Leica 20 x 0.40 LWD Hoffman Modulation contrast objective
specialized for 635 nm illumination
1280 × 1024 pixels, 3 pixels per μm, monochrome, 8-bit
< 0.5 s per image, using single 1W red LED (635 nm)
20 min cycle time for six slides, 5 min cycle possible with single slide
|
Tri-gas incubator
Temperature
Oxygen
CO2
Active air circulation
Volatile organic compounds
Particles
Disinfection
|
Integrated in instrument
30 °C to 45 °C ± 0.2 °C
5% to 20% ± 0.2 %
2% to 10% ± 0.2 %
Full purification/regeneration of gas volume every 10 minutes
Removed by Whatman Active Cap™ active carbon filter
Removed by HEPA filter which retains 99.97% particles >0.3 μm
Ozone-free UV illumination circulated air (254 nm)
|
Incubator control
Data acquisition
Network
Operating system
Data exchange
Data format for measurements
Data format for images
Monitor
Safety
|
Independent failure proof embedded system controls vital systems
Intel Core duo T2300E 1.66 Ghz 2Gb ram, fanless embedded PC
100Mb Ethernet
Microsoft Vista Pro®
Ethernet and front port for USB connection
Excel worksheets
JPEG
12.1” embedded touch screen 1024 × 600
Continuous internal system integrity check of separate subsystems of the instrument as an added safety feature.
|